* “美敦力”飛巡耳鼻喉科導航系統器械組回收警訊(2017/11/08)
*
*

許可證字號:衛署醫器輸字第022210

產品英文名稱:Medtronic Fusion ENT Navigation System Instruments

受影響規格/型號/批號:

名稱描述

型號

批號

直角電磁抽吸器

(Straight Suction, EM)

9733449

17022809, 17030108, 17022207, 17022106, 17020805, 17020904, 17021798, 17020987, 17011896, 17011221, 17011220, 16122237, 16122236, 16121661, 16121260, 16121259, 16112958

70度彎角電磁抽吸器

(70 Curved Suction, EM)

9733450

17030722, 17022421, 17022020, 17020619, 17011918, 17012417, 17010602, 16122724, 16122239, 16122038, 16120563, 16112362

90度彎角電磁抽吸器

(90 Curved Suction, EM)

9733451

17030728, 17022027, 17021426, 17013125, 17012724, 17011303, 16122767, 16122066, 16112965, 16112164

Fusion耳鼻喉科導航器械套組

(Fusion EMT Navigation Instrument Set)

9733452

包含上述(97334997334509733451)批號之套組

Fusion耳鼻喉科導航器械包

(Fusion EMT Navigation Instrument Kit)

9733908

包含上述(97334997334509733451)批號之套組

發布對象:醫療從業人員/醫療器材專業人員

警訊說明:(回收/矯正原因描述)

美敦力總公司獲知特定批次之受影響產品內之抽吸器,因係由具有磁特性(磁導率)之材料製成,導致對電磁(EM)導航能力干擾,故原廠進行特定批次產品回收。

國內矯正措施:

經查,國內並未進口受影響型號/批號產品,美敦力醫療產品股份有限公司對此警訊無須進行回收行動。

廠商聯絡資訊:

公司名稱:美敦力醫療產品股份有限公司
聯絡電話:02-21836168蔡小姐
聯絡人電子郵件:linda.tsai@medtronic.com
相關警訊來源(網址) 
  下載PDF檔案格式“美敦力”飛巡耳鼻喉科導航系統器械組回收警訊
* 資料來源- 衛生福利部食品藥物管理署
*
其他訊息
*
*
* “銳盟”鏈球菌檢測套組 (未滅菌)回收警訊(2018/2/5)
*
* “愛德華”印弗雷斯經皮膜鞘導引器組回收警訊(2018/2/5)
*
* “百靈列”手術導航系統安全警訊(2018/2/5)
*
* “希森美康”紅血球溶解劑(未滅菌)回收警訊(2018/2/5)
*
* “思威克”放射科病患用支架 (未滅菌)等3張許可證回收警訊(2018/2/5)
*
* “邁柯唯”心臟痲痺熱交換套管組等2張許可證安全警訊(2018/2/5)
*
* “漢尼帝克” 依利特自體輸血系統管套回收警訊(2018/2/5)
*
* “奧斯卡”暫時型心律調節器導線回收警訊(2018/2/5)
*
* “德塔斯”主動脈氣球導管幫浦安全警訊(2018/2/5)
*
* “邁柯唯”加熱器安全警訊(2018/2/5)